Store weight value (HOLD)
You can also continue displaying the weight determined once the load has been removed.
This makes it possible to put the baby down safely again before noting the weight:
– Press the HOLD key whilst the load is
on the scale.
The HOLD indicator flashes. The value
is stored until the scale switches off au-
tomatically.
– The HOLD function can be switched off
by pressing the HOLD key again.
This frozen display is a supplementary
device which does not have to be cali-
brated.
0. 0
appears in the display if the scale
has no load on it, otherwise the current
weight. The "HOLD" display vanishes.
Movements are damped by a special
damping system. However, if the baby
should move about on the scale too much,
the HOLD function is not activated and the
value displayed flashes. This prevents in-
accurate measurements.
6. Cleaning
Clean the tray and the housing as required with a domestic cleaning agent or commercially-
available disinfectant. Follow the manufacturer's instructions.
Under no circumstances use abrasive or acid cleaners, white spirit, benzene or the like
for cleaning. Such substances can damage the high-quality surfaces and the printing.
7. Transport
Before transporting the baby scale any distance, you should attach the transport securing
device. This protects the highly-accurate electronics.
Proceed as follows:
– Turn over the scale so that the base-
plate faces upwards.
– Attach the transport securing device
and push the spacers between the
frame and the tray support.
– Tighten up the screws on the base of
the scale.
Warning! Only tighten up the screws
by hand, do not use a tool!
– Turn the scale back up the right way.