Picture 7
To remove the introducer needle (C) hold the
cannula housing (G) firmly to your body. Place
one finger on the top of the cannula housing and
withdraw the introducer needle in a straight
manner using the introducer needle handle (A).
Remember to dispose the introducer needle (C)
safely (needle container). Discard your intro-
ducer needle according to local regulations. If
you inserted the infusion set all by yourself and
no safe needle container is at hand, consider
using the protective cover to cover the needle.
Ensure to avoid needle injury of others with the
risk of infection.
Picture 8
Attach the luer-lock (K) to the pump by tightening
it into the pump adapter. Follow the pump manu-
facturer's user manual to prime the transfer-set.
Prime the transfer-set until medication drips
from the connector needle tip. Do not leave air
bubbles in either the cartridge or the transfer-set.
Connect the transfer-set to the newly inserted
head-set and fill the empty space of the cannula
housing with a bolus of 1.0 units (U100 insulin).