Load fixed programs
Overwrite no. 90–99 ?
Yes = E
No = C
When loading the fixed programs, the programs 90– 99 will be
overwritten.
If a currently loaded program is in that range, it will be overwritten
as well.
The test program is displayed with the original data.

6 Create individual programs

6.1 Load program

As an example, load the free program 01 with the
button "R"
01.
010000000000000000000000c0020
0000
00
0
00
00 00 0
Note:
You have only 60 seconds to enter the individual firing para-
meters.

6.2 Enter preheating temperature

Press button "E". The cursor automatically selects the
preheating temperature. Use the numeric keypad 0 – 9 to
enter e.g. 500.
01000
0500 0000000000000c0020
0000
00
0
00
00 00 0

6.3 Enter heating rate

Press the cursor button " ". The cursor selects the heating
rate. Use the numeric keypad 0 – 9 to enter e.g. 80. The heat-
ing rate can be set to „max" (140 °C/min) with the button „C".
01000
0500
080 00000000c0020
0000
00
0
00
00 00 0
= next sub-menu
= previous sub-menu
E
= select sub-menu
M = back to main menu
Safety question when
loading fixed programs.
E
= Load and back to
main menu
C
= back to main menu
M = back to main menu
00,0 00,0 0
00,0 00,0 0
00,0 00,0 0

6.4 Enter firing temperature

Press the cursor button " ". The cursor selects the firing
temperature. Use the numeric keypad 0 – 9 to enter e.g. 940.
01000
0500
080
0000
00
0
00
00 00 0

6.5 Enter drying time

Press the cursor button " ". The cursor selects the drying
time. Use the numeric keypad 0 – 9 to enter e.g. 6.
01000
0500
080
0000
06
0
00
00 00 0

6.6 Enter preheating time

Press the cursor button " ". The cursor selects the preheating
time. Use the numeric keypad 0 – 9 to enter e.g. 3.
01000
0500
080
0000
06
0
03
00 00 0

6.7 Enter vacuum level

Press the cursor button " ". The cursor selects the vacuum
level. Use the numeric keypad 0 – 9 to enter e.g. 50.
01000
0500
080
0000
06
0
03
00

6.8 Enter vacuum time

Press the cursor button " ". The cursor selects the vacuum
time. Use the numeric keypad 0 – 9 to enter e.g. 1.0 (fixed
decimal point).
01000
0500
080
0000
06
0
03
00
Note:
The vacuum time must be at least 0.3 min shorter than the firing
time.
0
0940 0c0020
00,0 00,0 0
0
0940 0c0020
00,0 00,0 0
0
0940 0c0020
00,0 00,0 0
0
0940 0c0020
50
0
00,0 00,0 0
0
0940 0c0020
50
0
01,0 00,0 0
33