Cleaning and Maintenance
Cleaning
All surfaces of the unit allow for cleaning with a damp cloth and if required a mild soap
lotion.
Attention:
Damage to the surfaces
All surfaces can be damaged by improper cleaning.
Never use:
▪ Chlorine bleach or any cleaning agent including chlorine
▪ Ammonia
▪ Abrasive cleaning agents such as cleaning rags, scrubbing agents or
any other agents which include metal components
As a result of vibrations generated during operation the device might
slide off the table top.
To ensure a stable position clean surface and feet with a damp cloth
and ethanol regularly.
Maintenance
The unit is maintenance-free. Any necessary repair must be performed by an authorized
Heidolph distributor.
Please contact Heidolph Instruments or your local Heidolph Instruments distributor.
Possible errors and how to resolve
Malfunctions
On-/Off switch does not
enlighten
Shaking movement stops
If you experience a malfunction which can not be resolved, please contact your authorized
Heidolph Instruments distributor immediately.
Cleaning and Maintenance
Troubleshooting
Possible reason
No power
Light diode defective
Pre-set timer elapsed
Thermal motor circuit
breaker triggered by motor
overload.
▪ Motor defective
▪ Control electronics
defective
15
Troubleshooting
▪ Check size and shape of
plug and compatibility to
your electrical socket
▪ Check circuit breakers
Contact your local Heidolph
Instruments distributor
Restart timer, if necessary
Wait 20 minutes, decrease
overall load
Contact your local Heidolph
Instruments distributor