POWER SUPPLY
Open the battery compartment cover and insert 2 x 1,5V AAA alkaline
batteries (ensure correct polarity).
IMPORTANT
Calibration of inclinometer:
- before first use
- before important measurements
- after a hit or drop
- after a large fluctuation in temperature
HOW TO CALIBRATE
Lay down the unit on a flat surface and switch it on.
Press the buttons ON/OFF and C simultaneously; the display shows --1--.
Press the button ON/OFF again; --1-- is blinking.
Then --2-- appears in the display.
Now turn the instrument by 180°.
Press the button ON/OFF again; --2-- is blinking shortly.
Now the unit is calibrated to absolute 0 and the currently measured value is
displayed.