321
3
Adjustments
3.1
Before getting started
3.1.1
Table of specifications
In carrying out the following adjustments, it is necessary to know the specifications of the balance in
service. You may download the specifications for servicing from our website:
http://www.precisa.com
The following instructions refer often to that specifications, have a printout of them always at
disposal!
3.1.2
Placing loads
Place all loads always carefully and, unless otherwise asked, in the center of the weighing pan!
3.2
Setting the balance to the factory mode (the balance is open)
1.
Remove the jumper J1 from the main board.
2.
Level the balance with the levelling screws and connect it to the mains with the power adapter.
The balance starts up in the factory mode.
Adjustments
Pin header
for jumper J1
42