• Calibration with manual values
Example
two-point calibration pH 6.79 e pH 4.65 (DIN19267)
:
Access the Setup menu for pH and select in P1.1→ Custom, press twice the key
measurement and position in pH mode
Press
to enter the calibration mode.
Rinse the electrode with distilled water and gently dab with paper towel. Dip the electrode in the first
pH buffer solution (e.g. pH 6.79).
Wait for the pH value on the display to stabilize; when the icon
modify it using the directional keys by entering the correct one (e.g. pH 6.79), as suggested by the
string "ADJUST THE VALUE" and by the icon
Note: Check the buffer value according to the temperature
When the icon
appears again, press the key
value flashes on the display and the beaker icon appears with the buffer identification colour
Remove the electrode, rinse with distilled water, dab gently with paper towel and dip it in the next
buffer (e.g. pH 4.65).
Wait for the pH value on the display to stabilize; when the icon
flashes, modify it using the directional keys by entering the correct one (e.g. pH 4.65), as suggested by
the string "ADJUST THE VALUE" and by the icon
When the icon
measured value flashes on the display, the Slope% and, next to the first beaker, the icon with identifying
colour of the second buffer
At the end of the second calibration point, the instrument automatically returns to measure mode.
To perform a one-point calibration just press the key
Note: If you are working with manual temperature compensation (MTC), update the value before
calibrating the instrument.
• Performing pH measurement
In measure mode, press the key
Connect the electrode to the pH / ORP BNC of the instrument (green).
If the user does not use an electrode with a built-in temperature probe or an external probe NTC 30KΩ,
it is recommended to manually update the temperature value (MTC).
Remove the electrode from its tube, rinse with distilled water and dab gently with paper towel.
Check the presence and eliminate any air bubbles in the membrane bulb by stirring vertically (as for
the clinical thermometer). If present, open the side cap.
Dip the electrode in the sample, while keeping it slightly stirred.
The scrolling on the display of four red bands
Consider the measurement truthful only when the stability icon appears
UM Serie 50VioLab EN rev.2 26.07.2021
.
appears again, press the key
appears.
and move to the pH parameter indicated by the icon
18
appears and the value flashes,
.
to confirm the first point; the actual measured
appears and the actual value
.
to confirm the second point; the actual
after finishing the first point.