Messgaskühler / Sample Gas Coolers EGK 1/2
6.2 Trouble shooting
Problem / Failure
No display
Display blinks with
- High Temperature
- Low temperature
Display of an error
Error 01
Error 02
Condensate in gas outlet
Reduced gas flow
6.3 Replacement of the peristaltic pump's hose (just in case pump
is fitted)
Disconnect input and output hoses (Take care of the safety instructions!)
Loosen but don't remove the centre knurled thumb screw. Push screw downwards.
Remove cover
Push terminals side wards.
Replace hose and step backwards the above steps for mounting.
26
Possible cause
- no power
- fuse blown
- operational temperature not yet
reached
- cooling capacity to low, even
though cooler is running
- gas flow / dew point / gas tempe-
rature to high
- fan broken
- controller defect
- broken wire
- short circuit
- condensate vessel full
- valve in automatic condensate
drain is stuck
- cooler overloaded
- clogged gas path
- condensate outlet clogged by ice
BX450001, 01/2012
- check power supply
- Check fuse and change it if
necessary
- wait for 20 minutes maximum
- make sure that air can circulate
free and that ventilation louvres
are not obstructed
- check application parameters,
install pre-separator
- check fan, replace if necessary
- send cooler for inspection
- Temperature sensor defect: send
cooler for repair
- Temperature sensor defect: send
cooler for repair
- drain vessel
- flush both directions
- check limiting parameters
- check / flush heat exchanger
- send cooler for inspection
Art. Nr. 90 31 059
Solution