6. Placement of Restoration
Place restoration. Apply pressure and then release to allow the hydraulic
pressure to dissipate. Repeat this procedure at least one more time and
ensure that the restoration is completely seated.
Remove as much excess as possible before the compomer is light-cured.
If any excess compomer remains after curing, remove with curette, scalers
or finishing burs/discs or strips.
7. Light-Curing
Expose all surfaces of the restorations to the dental curing light for 40
seconds each. Depending on the number of surfaces of the restoration,
the diameter of the curing light outlet and the size of the tooth, this results
in up to 5 curing cycles per restoration.
10
8. Occlusal Adjustments and Finishing
Make occlusal adjustments and finish the margins with finishing burs.
Remove excess from proximal surfaces with diamond strips or with the
diamond tips for the EVA contra-angle handpiece.
Use polishing discs (Enhance) on inlay margins where accessible. Use
polishing pastes (Prisma Gloss and Prisma Gloss Extrafine) additionally
or alternatively. Work interdentally with aluminium oxide strips.
STORAGE
The Prime&Bond NT bottle and Dyract flow syringes should be tightly
closed immediately after use.
Keep out of sunlight.
Keep Prime&Bond NT in a well ventilated place.
To be stored at temperatures between 10 °C and 24 °C.