Verfügbare Sprachen

Range
Materials (external antenna)
11.1
Configuring the Internet protocol of the mobile terminal
NOTICE
If the WLAN connection is lost during the configuration, settings made may be lost.
Make sure that the WLAN connection is not disconnected while configuring the device.
NOTICE
In principle, avoid simultaneous access to the measuring device via the service interface
(CDI-RJ45) and the WLAN interface from the same mobile terminal. This could cause a
network conflict.
Only activate one service interface (CDI-RJ45 service interface or WLAN interface).
If simultaneous communication is required, configure different IP address ranges, e.g..
192.168.1.212 (CDI-RJ45 service interface).
11.2
Preparing the mobile terminal
Enable WLAN reception on the mobile terminal.
11.3
Establishing a connection from the mobile terminal to the meas-
uring device
1.
In the WLAN settings of the mobile terminal:
 Select measuring device using SSID (e.g. EH_Promass_300_A802000).
2.
If necessary, select the WPA2 encryption method.
3.
Enter the password: serial number of the measuring device ex-works (e.g.
L100A802000).
 LED on display module is flashing: it is now possible to operate the measuring
device with the web browser, FieldCare or DeviceCare.
The serial number can be found on the nameplate.
To ensure the safe and swift assignment of the WLAN network to the measuring point,
it is advisable to change the SSID name. The new SSID name should be uniquely assigned
to the measuring point (e.g. device tag), since it is displayed as a WLAN network.
Endress+Hauser
• Internal antenna: typically 10 m (32 ft)
• External antenna: typically 50 m (164 ft)
• Antenna: ASA plastic (acrylonitrile styrene acrylate) and nickel-plated
brass
• Adapter: Stainless steel and nickel-plated brass
• Cable: Polyethylene
• Plug: Nickel-plated brass
• Angle bracket: Stainless steel
EA01158D