Herunterladen Diese Seite drucken

Seca 286 Bedienungsanleitung Und Garantieerklärung Seite 73

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Fitting the ultrasound head
1. Screw the long threaded rod hand-tight into the connecting nut of the
second column element.
2. Push the third column element over the threaded rod onto the column
connector of the second column element.
The ultrasound head is fitted to the third column element and connected to
the modular cable of the multifunctional display.
ATTENTION!
Device damage as a result of assembly error
The third column element is loose on the second column element.
Hold the third column element firmly as you lay the device on the
floor.
Ensure that the third column element is in the right position before
you fit the ultrasound head.
ATTENTION!
Malfunction as a result of faulty multifunctional display
When the device is laid down, the multifunctional display is directly on
the floor and may be damaged.
Lay the device down slowly and carefully on a soft surface, a
blanket, for example.
1. Tilt the device and carefully lay it on the floor.
2. Push the ultrasound head onto the free end of the threaded rod.
3. Put the plain washer onto the threaded rod.
4. Screw the connecting nut onto the threaded rod.
5. Tighten the connecting nut.
ATTENTION!
Damage to device and malfunction as a result of assembly error
If the connecting nut in the ultrasound head is not tightened enough,
the column elements will be too loosely connected. The device will be
unstable and faulty measurements will occur.
Tighten up the connecting nut so that the columns are firmly located
on one another and there is no gap between the column connector
and the third column element.
Before you really get started ... • 73
English

Werbung

loading

Diese Anleitung auch für:

287