MACHEREY-NAGEL NANOCOLOR VARIO HC Handbuch Seite 18

Water analysis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Display
01-11-2011 12:15:00
Press
surrounding light conditions.
Sound
Press
The icon
in the display shows the adjusted sound level.
T–Set
Press
to select the calibration mode.
The calibration mode allows you to calibrate all programmed temperatures.
After calibration, press
Press
to select the test mode.
The test mode allows you to test all programmed temperatures, the calibration is not
changed.
01-11-2011 12:15:00
Press
to restore factory settings.
Press
to start the program.
Deviations in temperature are displayed in different colors:
Deviation
< 1 °C
1 < x < 2 °C
> 2 °C
Press
to save the calibration or test protocol on a USB stick.
18
to adjust brightness
to adjust the sound level.
to confirm and save the new calibration data.
Color
green
yellow
red
, contrast
and saturation
to the
www.mn-net.com