Section 4 Operation
Setup
CAUTION
CAUTION
Start Up
4-2
Refrigerated baths should be left in an upright position at room
temperature (~25°C) for 24 hours before starting. This will en-
sure the lubrication oil has drained back into the compressor.
Before starting the circulator, double check all USB (optional),
electrical and plumbing connections.
Do not run the circulator until fluid is added to the bath. Have
extra fluid on hand. If the circulator will not start refer to Section 6
Troubleshooting.
• Place the circuit protector located on the back of
the circulator to the I position.
The circulator will display
• Press
• Ensure the start symbol has a highlight box around it, if not
use the arrow keys to navigate to the symbol.
20.0°C
Highlighted
Start Symbol
• Press
into a stop symbol (
Stop Symbol
Note It will take 30 seconds for the compressor to start.
Note After start, check external plumbing connections for leaks.
The SP1 and Menu portions on the top of the display are used to
view and/or change the circulator's settings. They are explained in
detail later in this Section.
System Standby
, the Start Display appears.
SP1
Menu
24.2°C
Start Display
. The circulator starts and the start symbol turns
).
.
Reservoir
In
Fluid
Temperature
20.0°C