Figure 3-2 Location of the Zero Button.
3-2
All manuals and user guides at all-guides.com
After the Brooks Quantim has been fully installed, you must perform the
zeroing procedure. This procedure ensures that the instrument responds
properly to zero flow condition and sets a baseline for flow. To perform the
zeroing operation on the meter/controller, use the ZERO button, which is
located on top right hand side (outlet side) of the instrument. Zero button
must be pressed for at least 3 seconds to initiate zeroing procedure. The
status LED will flash red on and off during the zeroing procedure. (See Figure
3-2).
Zero Procedure
1.
Prepare the instrument for zeroing:
Installation and Operation Manual
Part Number: 541B029AAG
X-CM-QmB-eng
June, 2017