6. Plug the mains cable into the socket on the back of the instrument.
7. Switch on the instrument. Symbols on or near the power switch of the unit have the following
meanings:
replacemeNt caBle
Should the mains lead need replacement, a cable of 1mm 2 of harmonized code H05VV-F connected
to an IEC320 plug should be used. If IN douBt coNsult a QualIfIed electrIcIaN.
eNVIroNmeNtal coNdItIoNs
The CB-200D-IB is designed to operate under the following conditions:
• Indoor use
• Ambient temperature range +5ºC to +40ºC
• Altitude to 2000m
• Relative humidity not exceeding 80%
• Mains supply fluctuations not exceeding 10%
• Over voltage category II IEC 60364-4-443
• Pollution degree 2
Note: The control specifications are quoted at an ambient temperature of 20ºC. The specification may
deteriorate outside an ambient temperature of between 10ºC and 30ºC.
The instrument has been tested for radio frequency interference and is certified under EN61326
I
Mains Switch On
o Mains Switch Off
7