Operation
Allowed test gas concentration:
Ex channel of the
infrared sensor
CO
channel of the
2
infrared sensor
Ex:
O
2
CO:
H
S:
2
NO
2
Test gas concentration of other gases:
see instructions for use of the respective DrägerSensors.
1)
Depending on the selected data set.
2)
Depending on the measuring range and the measuring accuracy.
1
Connect the test gas cylinder with the calibration cradle.
2
Vent the test gas into a fume cupboard or into the open air (with a hose
connected to the second connector of the calibration cradle).
!
CAUTION
Risk to health! Never inhale the test gas.
Observe the hazard warnings of the relevant Safety Data Sheets.
3
Switch on the instrument and insert it into the calibration cradle.
Press the [ + ] key and keep it pressed for 5 seconds to open the calibration
4
menu, enter the password (password on delivery = 001).
Use the [ + ] key to select the single-gas adjustment function. The symbol
5
for span adjustment »
6
Press the [OK] key to start the channel selection.
In the case of the CO2 channel there is a 2-stage calibration routine:
The zero adjustment takes place first and is then followed by the span
adjustment.
The display flashes the gas of the first measuring channel, e.g. » ch4 - %LEL «.
7
Press the [OK] key to start the adjustment function of this measuring
channel, or use the [ + ] key to select another measuring channel
(O
- vol. %, H
2
The test gas concentration is displayed.
46
20 to 100 %LEL
0.05 to 5 vol. %
40 to 100 %LEL
10 to 25 vol. %
20 to 999 ppm
5 to 99 ppm
5 to 99 ppm
« flashes.
S - ppm, CO - ppm, etc.).
2
1)2)
/ 5 to 100 vol. %
2)
1)2)