Herunterladen Inhalt Inhalt Diese Seite drucken

Limits Of Use Of Product; Proper Use - Pfeiffer Vacuum HIPACE 80 NEO Betriebsanleitung

Turbopumpe
Vorschau ausblenden Andere Handbücher für HIPACE 80 NEO:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

► Before working on the high vacuum connection, wait until the rotor has stopped completely (rota-
► Never put the device into operation with the high vacuum connection open.
► Keep lines and cables away from hot surfaces (> 70°C).
► Never fill or operate the unit with cleaning agents or cleaning agent residues.
► Do not carry out your own conversions or modifications on the unit.
► Observe the unit protection class prior to installation or operation in other environments.
2.4 Limits of use of product
Installation location
Air pressure
Installation altitude
Rel. air humidity
Protection class
Overvoltage category
Permissible protection degree
Degree of pollution
Ambient temperature
Permissible surrounding magnetic field for air cooling
Permissible surrounding magnetic field for water cooling
Maximum irradiated thermal output for air cooling
Maximum irradiated thermal output for water cooling
Maximum permissible rotor temperature of the turbopump
Maximum permissible bakeout temperature at the high
vacuum flange
Tbl. 3:
2.5 Proper use
► Use the turbopump only for generating vacuum.
► Use the turbopump only in combination with a suitable backing pump that can deliver up to the
► Use the turbopump only in closed indoor areas.
► Use the turbopump only for the evacuation of dry and inert gases.
tion speed f = 0).
Permissible ambient conditions
Notes on ambient conditions
The specified permissible ambient temperatures apply to operation of the turbopump at
maximum permissible backing pressure or at maximum gas throughput, depending on the
cooling type. The turbopump is intrinsically safe thanks to redundant temperature monitor-
ing.
● The reduction in backing pressure or gas throughput permits operation of the turbo-
pump at higher ambient temperatures.
● If the maximum permissible operating temperature of the turbopump is exceeded, the
electronic drive unit first reduces the drive output and then switches it off where neces-
sary.
required maximum fore-vacuum pressure.
Weatherproof (internal space)
530 hPa to 1060 hPa
Max. 5000 m
max. 80%, at T <31°C,
up to max. 50% at T <40°C
III
II
IP54,
Type 12 according to UL 50E
2
5 °C to 30 °C with convection cooling
without gas throughput
5°C to 35°C with air cooling
5°C to 40°C with water cooling
3.1 mT
3.7 mT
0.9 W
1.2 W
90 ℃
120 ℃
Safety
79/130

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis