Continuity Test
With this function you can check electrical circuits, cables or
components for continuity.
Attention!
Do not perform continuity tests on voltage-carrying components
or circuitry.
Measuring Procedure
1. Make sure the component or circuit you intend to measure is
voltage free.
2. Set the rotary function switch (3) to 2000 Ω
3. Connect the black measuring cable to the COM input (6)
and the red measuring cable to the V/Ω/
connect the measuring tips to the component of circuitry.
4. An acoustic signal sounds if the resistance is lower than
30 Ω, confirming that the component or the circuit has
passed the continuity test.
Cleaning and Storing
Only clean the device with a soft moist cloth and allow it to
dry completely before the next use.
Do not use abrasives and aggressive chemicals.
Store the unit in a dry place.