4.2

Display overview

1 Alarm display (if an alarm is triggered)
2 Displays which gas inlet is being measured: MEAS (gas
inlet measurement) / REF (gas inlet reference)
3 Measured value
4 Display response factor
4.3

Select the instrument menu

NOTICE
► The navigation bar automatically disappears after
120 seconds of inactivity.
► The menu cannot be called up while the UV lamp is being
checked (see chapter 5). The device information is still
displayed.
1. Use a magnetic pen to tap any point of the display so that
the navigation bar appears.
2. Select "menu".
3. Confirm "WARNING" with "ok".
4. Enter the password. The password is "0002". Set the
numbers by tapping on "more" or "less" and switch to the
next number item with "next". Once the password is
correctly entered, confirm with "accept".
WARNING
Danger due to interruption
The measuring function is interrupted when accessing the
instrument menu. No measurement takes place, all alarms
are disabled and the current loop, the relay and the LEDs are
switched to maintenance mode. The measuring function
resumes when leaving the menu.
5. Confirm the "confirm" screen with "ok". This accesses the
instrument menu.
4.3.1

Response factor

A response factor must be entered in the PID 5000 based on
the calibration with isobutylene gas. This factor reflects the
sensitivity of the known volatile organic compound compared
to the calibration gas and can be set between 0.10 and 10.00
in steps of 0.01. The response factor depends on the oil type
to be measured, as different oils are stimulated at different
levels by the UV lamp.