InPro 6800 G / 6850i G / 6900i G / 6950i G O
2.8.2 Rated data
For rated data please observe section 2.7.2
2.8.3 Special conditions
In type of protection intrinsically safe apparatus, the
analytical probes (InPro 6000 Series with InFit 76X/Y
and InTrac 7XX/YY Series housings) equipment is des-
ignated with the following specific conditions of use.
1. The oxygen sensor shall be installed in compli-
ance with the enclosure, mounting, spacing and
segregation requirements of the ultimate appli-
cation, including a tool removable cover, and is
suitable for use with FM Approved InFit 76X/Y and
InTrac 7XX/YY Series Housings.
2. Process temperature no greater than +130 ºC.
3. Maximum permissible working pressure is
12 barg (174 psig).
4. Potential Electrostatic Charging Hazard –
To prevent the risk of electrostatic sparking, the
non-metallic surface should only be cleaned with
a damp cloth.
5. Enclosures containing titanium constitute a poten-
tial risk of ignition by impact or friction. Care must
be taken into account during installation and use
to prevent impact or friction.
© 03 / 23 METTLER TOLEDO
Printed in Switzerland
12 mm
2
InPro 6000 G Series
52 206 477 F
13