10.0
Technical specifications
Air flow rate of pump
Max. vacuum
Vacuum readout
Additional air regulation
Collection jar
Suction hose
Voltage
Current input (max.)
Power consumption
Power cable
Operating time
Fuse
Protective earth conductor resistance
Earth leakage current
Enclosure leakage current
Patient leakage current
Heat emission
Noise level
Ambient conditions
Transport/storage
Operation
Dimensions HxBxT
Weight
Protection class (EN 60601-1)
Degree of protection
Protection category
Classification acc. to Annex IXEEC
directions 93/42/EEC
CE marking
Rules applied
UMDNS-Code
Reference-No.
Canadian Classification
Device Group
PNC
Risk Class
Description
* 1 bar ≈ 750,06 mm Hg ≈ 1000 hPa / depends on daily atmospheric pressure
18
16 ± 2 l/min
-80 kPa*
-1...0 bar (± 25 mbar) (mm Hg; kPa)*
mechanical regulating valve
1 l jar made of glassor
1.25 l plastic (TPX)
or 1.5 l Receptal®-system
ø 6 mm, 1.30 m length
230 V~ (+6%/-10%) 50/60 Hz
Special voltage:
115 V~ (+6%/-10%) 50/60 Hz
approx. 0.7 A for 230 V~
approx. 0.95 A for 115 V~
approx. 75 VA
2 m
Interrupted use over approx. 45 min.,
cooling period approx. 60 min.,
depending on ambient temperature
slow-blow 1,25 A/H for 230 V~and 115 V~
—
—
N.C. < 0,1 mA
—
65 J/s
53,4 dB (A) @ 1m (acc. to ISO 7779)
-30...+50°C20...80 % humidity,
non-condensingair pressure 700...1060 hPa
+5...+35°C20...80 % humidity,
non-condensingair pressure 700...1060 hPa
206 x 362 x 174 mm
3.5 kg
II
Type B
IP 00
IIa
CE 0124
EN 60601-1: 1990 +0A1: 1993 + A2: 1995
EN 60601-1-2: 1993 (EMV / EMC)
EN 30993: 1994
10-219
320.0100.0 (with 1 l jar made of glass)
320.0101.0 (with 1,25 l TPX-jar)
320.0102.0 (with 1,5 l Receptal®-system)
Ear, Nose and Throat
77QBW
2
ASPIRATOR, TRACHEAL
English
Issue of Technical Specifications: 21.06.2004