Herunterladen Inhalt Inhalt Diese Seite drucken

mectron Compact Piezo LED Installationsanleitung Seite 44

Vorschau ausblenden Andere Handbücher für Compact Piezo LED:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 97
COMPACT PIEZO LED
Problem
No liquid flows out from the
insert during operation
Poor performance
Table 9 – Quick Troubleshooting
36
Possible Cause
The insert is the type without
a liquid passage
The Water function is not
active
The insert is clogged
The handpiece is clogged
The insert is not correctly
tightened on the handpiece
Insert broken, worn or
deformed
Solution
Use an insert type with a
liquid passage
Activate the Water function
Unscrew the insert from the
handpiece and release the
water passage of the insert
by blowing compressed air
through it.
If the problem persists,
replace the insert with a new
one
Contact an Authorised Mec-
tron Service Centre
Unscrew the insert and screw
it back in correctly using the
Mectron torque wrench (See
Chapter 4.4 on page 14
of the Use and Maintenance
Manual)
Replace the insert with a
new one

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis