Operation
The Air Guard can only be reset by purging the system (see
chapter 4.4.8).
Open the alarm overview on the Air Guard touchscreen to see
which alarm was exceeded (see chapter 4.4.4). The
corresponding measured values flash yellow (warning) or red
(alarm or malfunction).
If the flow rate is too low, the symbol for the differential
pressure sensor turns red (see chapter 4.4.4). If the flow rate
is sufficient once again, the symbol changes and the Air Guard
can be reset by purging.
PID 5000 lamp test
During the lamp test, the "CHECK" message appears on the
PID 5000 display. The functionality of the UV lamp is checked.
This process takes approx. 100 seconds and is triggered after
restarting the Air Guard, after the "FLOW ERROR" error and
cyclically every 7 days (operating time of the lamp since the
last lamp test). It is not possible to carry out a measurement
during this time (see chapter 4.3.3). No operation is required
during this lamp test. After the lamp test has been completed,
the PID 5000 automatically switches to measuring mode. The
lamp test is stored in the messages for follow-up purposes
(see chapter 4.4.7).
4.4

Touchscreen operation

The Air Guard is operated via the touchscreen. The user can
navigate from each menu element directly to the
corresponding menu using the four operating keys below the
touchscreen.
The keys below the touchscreen are assigned as follows:
F1

Overall view

This key opens the "Overall view" menu (see
chapter 4.4.4).
F2
Purging
This key opens the "Purging" menu (see chapter
4.4.8).
F3
Acknowledge alarm tone.
The audible alarm is stopped. The visual alarm
and the potential-free contacts are not influenced
by this action.
F4
Settings
This key opens the "Settings" menu (see chapter
4.4.7).
4.4.1

General

The Air Guard can be operated without signed-in users. A user
must be logged in to purge the Air Guard or to change settings.
NOTE
The "Service" menu item cannot be selected as it is only
enabled for service personnel.
Main menu
Press the 3 bars
to bring up the main
menu. Using the main menu it is possible
to navigate away from any screen.
24
Symbols
Various symbols appear in the upper area; their function and
meaning are explained in the following.
No user is logged in.
A user is logged in.
The device is purging.
There is a new alarm pending. Clicking on the symbol
opens the "Messages" screen (see chapter 4.4.7).
The symbol is displayed as soon as the Standby mode
is marked.
4.4.2

Progress screen

The progress screen is
displayed during the
warm-up phase of the
sensors and the system
start-up.
4.4.3

Language selection

The language selection
screen appears as
soon as the start-up
phase is complete.
Choose the desired
language.
4.4.4
Overall view
The overall view screen
displays all measured
values of the installed
sensors. It also displays
the status of the
pressure switch.
The various display
variants of the
individual readings are listed below:
Measured value from relevant sensor is in normal
range.
Measured value from relevant sensor is in
warning range.
Measured value from relevant sensor is in alarm
range.
Sensor has a malfunction. No measured value is
displayed, only #.
Measured value of flow rate at pressure sensor in
normal range.