Herunterladen Inhalt Inhalt Diese Seite drucken

HBM CMA Bedienungsanleitung Seite 27

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Template: High Level Voltage
Parameters
Calibration period
(days)
Measurement
location ID
1)
Typical values for the CFT/20kN
force transducer
Other templates, such as the HBM "Signal Conditioning" template, can also be
created.
Tip
More information can be found on the HBM Internet site www.hbm.com/
HBMdoc, under TEDS.
8.2
Replacing measuring chain components
When replacement is necessary, in principle, the piezoelectric transducer and
the charge amplifier or connecting cable can be replaced. But the following
always applies: components must only ever be replaced with components of
the same type, e.g. a CFT/5kN force transducer only ever for a CFT/5kN, and
a CMA39 charge amplifier for a CMA39. This ensures that the transducer and
charge amplifier have the same nominal (rated) force and sensitivity.
Replacement of the connecting cable is non‐critical.
Charge amplifiers are adjusted individually. This ensures that when the
amplifier is replaced, the series variation is less than "0.5%.
CMA
1)
Value
Unit
0
days
0
A02559_04_YI0_01 HBM: public
Rights
Explanation
level
CAL
Time before recalibration,
calculated from the date
specified under Calibration
Date.
USR
Identification number for the
measuring point (assigned
by the user). Possible
values: 0 to 2047 (numbers
only). The HBM Channel
Comment template is also
available as a supplement.
Start‐up
25

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis