7. Most stations have a temperature indicator or heater lamp to
show that the station has reached the set temperature. It is better
to measure until the temperature stabilizes.
8. Use K type sensor (TP-01) to measure the solder temperature of
soldering pot or other liquid temperature.
9. If the LCD displays "1" which means the sensor is burned out.
Please replace the sensor with a new one.
10. If the LCD displays "B" which means the battery is dead, replace
the battery with a new one.
Appendix A: thermometer use
1) The thermometer designs for measuring the temperature.
2) Soldering stations should be calibrated when the heating element
or sensor on the soldering iron has been replaced and at the
manufacturer's recommended specified time intervals as the
electronics within the system has drifted over time.
3) Operators may, at their discretion, calibrate a soldering station
when the tip is exchanged with the one of significantly different
specifications. (E.g. removing a very small tip and replacing it
with a very large tip).
4) The ratio of surface area of tip will have an effect on the tip
temperature. To ensure the same temperature, the station needs
to be calibrated after changing a new tip.