en - Areas of application
30
The APFD 200 insert for reprocessing air regulators must be supplied
with pressure for reprocessing. The pressure ensures that wash water
cannot enter the air regulators during reprocessing, assuming the
correct adapters are used.
The compressed air must correspond to the quality of breathing air
as a minimum. For this, observe the national and regional regula-
tions.
In addition, a flexible compressed air connection must be provided
on-site. The compressed air connection must be equipped with one
of the following quick-release couplings:
- CEJN series 341
- Parker Rectus series 25S
Colour coding
The small parts of a dismantled air regulator are reprocessed in a
compartment of the small-part basket. Colour-code the compart-
ments and the ring holders or the load item carriers to ensure that the
small parts can be assigned to the right air regulators after repro-
cessing.
Coloured tags (2 in each colour) are provided with the inserts for this
purpose.
For each colour, attach one tag to the ring holders or the load item
carriers and one tag to the small-part basket compartments.
When using other systems for identification, ensure that these are
suitably chemical-resistant and temperature-resistant.