Operation
Max. drain
distillate:
Additional evap-
oration time
Forced draining
of the collector
Standby timer
after educt
container runs
empty

Parameters Hei-VAP

Parameters Hei-VAP
In the sub-menu
Hei-VAP can be stored (vacuum, hysteresis, rotational speed, heating bath temperature,
recirculation chiller temperature).
These values are for informational purposes only and may be included in a new recipe.
Hei-VAP parameters
Touch the
to open the submenu and to enter the required values for the connected rotary evapo-
rator by hand.

Data logging

See section "Data recording" on page 118.
Selectable if a specific number of distillation drainage steps
is specified in the start window:
In the example: drain distillate six times, then the optional
additional evaporation time runs. This allows a desired
concentration to be achieved!
max. filling: Add. evaporation
The parameters
time
max. drain distillate: additional
and
evaporation time
be activated at the same time!
Forced draining of the collector active: Collector will be
drained after 1 – 9999 seconds, regardless of the filling
level.
Range: 1 – 9999 s.
Activate/deactivate forced draining of the collector (deter-
mines previous parameter!).
Range: active / inactive
Time until the standby mode is activated after the educt
container runs empty.
Range: 1 – 9999 s.