Electrocution and fire hazard. Attach the power cord so that it cannot touch hot surfaces.
A PC with TEIS2 software installed is necessary to operate the analyzer.
1. Put the PC near to the analyzer. Make all of the necessary PC connections.
2. Use the supplied USB cable to connect the PC to the analyzer. Refer to
Note: Make sure that the USB cable is not parallel to a power cable or electrical noise interference on the
connection can occur.
3. Make sure that the power switch of the analyzer is set to OFF. Connect the supplied power cord
to a grounded electrical outlet.
Figure 8 Connect the PC and power

4.9 Install the software

Note: Do not start the TEIS2 software at this moment. Not loading the correct database will cause the instruments
to malfunction.
Use the supplied USB flash drive to install the TEIS2 software on the PC as follows.
1. Connect the USB flash drive to the PC.
2. Go to USB > Software > 3rd part. Install two software programs:
a. AbdeRdr90_en-Us.exe
b. vcredist_x86.exe (internet version) or dotNetFx40_Full_x86_x64.exe (off-line version)
3. Go to USB > Software > TEIS > V2.x.x.
4. Install Setup_TEIS2.msi.
5. Open the TEIS2 software. "No database found" shows.
14 English