g.
1 x

5.3 revPIP mode

Program especially for pipetting liquids with high viscosity, high vapor pressure, or foaming media.
Volume and speed setting, see Set volume, p. 54 and Set the aspiration and dispensing speed, p. 55.
a.
3 x
b.
c.
1 x
d.
1 x
Transferpette® -8/-12 electronic 997200
g.
End pipetting: Pressing the pipetting key once dis-
penses the liquid and triggers the blow-out. After dis-
pensing the residual liquid (blow-out), the display
jumps back to the set mode (start position).
The display shows a maximum of 19 cycles.
a.
Call up menu selection: Pressing the menu key three
times takes you to the program selection.
ð
'Mode' flashes.
b.
set revPIP mode: Use one of the arrow keys to scroll
through the modes until 'revPIP' appears.
ð
'Mode' continues to flash.
c.
Confirm revPIP mode: Press 'Enter' key.
ð
The display now shows 'blo' for blow-out.
d.
Preparing for pipetting: Press the pipetting key once to
move the piston to its start position.
ð
The arrow in the display points upwards (aspiration).
5 Pipetting programs