Herunterladen Inhalt Inhalt Diese Seite drucken

Signal Outputs - Bühler technologies TC-Standard Serie Kurzanleitung

Peltier-messgaskühler
Vorschau ausblenden Andere Handbücher für TC-Standard Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
TC-Standard (+)

4.4 Signal outputs

The device provides different status signals. The maximum
switching load of the alarm outputs is limited to
250 V AC/150 V DC, 2 A, 50 VA each.
An alarm is triggered by the alarm contact/status output (S2)
if the temperature of the cooler is outside the specified limits.
It does not indicate if the alarm was triggered due to excess
temperature or insufficient temperature.
The front film has three LEDs:
Colour
Marking
Function
Red
S2
High/low temperature, device error
Yellow
S1
---
Green
OP
Normal operation
The LEDs OP and S2 indicate the device status similar to S2.
If the option "temperature signal" is built in, the unit has a
signal output via the analogue output to indicate the actual
cooler temperature.
When the moisture detector (optional) is installed, an alarm is
activated by the alarm contact/status output (S2) if the mois-
ture is still present in the prepared sample gas. Thereby, no
distinction is made between the alarm/cable break triggered
by moisture detector 1 or 2. This information is displayed by
an error message instead.
The temperature signal can be read via the panel plug (S3) us-
ing the M12x1 connector. This plug is located next to the mois-
ture detector connectors at the top of the cooler.
Description of signal outputs
Function /
Description
contact type
Re-
internal
the following
gard-
changeover
device statuses
ing S2)
contact:
can be indic-
max.
ated via two
250 V AC /
switching out-
150 V DC, 2 A,
puts:
50 VA
BX440026 ◦ 12/2020
Contact between 3 and
2 closed (alarm)
• No mains voltage
and/or actual tem-
perature outside
the alarm
thresholds
Contact between 3 and 1
closed (ok)
• Mains voltage at-
tached + actual
temperature within
the alarm
thresholds
With moisture detector
option
Contact between 3 and
2 closed (alarm)
• The moisture de-
tector registers re-
sidual humidity in
the sample gas or
cable break: Error
message
Contact between 1 and 3
closed (ok)
Re-
4-20 mA
Signalling of
gard-
analogue
actual temper-
ing S3)
output
ature (please
(R
 <500 Ω)
use shielded
Load
cables)
5 Operation and control
NOTICE
The device must not be operated beyond its specifications.
After switching on the cooler the block temperature will be
displayed. The display will flash until the block temperature
has reached the preset target value (± adjustable alarm
range). The status contact is in the Alarm position.
Once the target temperature range has been reached, the
temperature will continuously be displayed and the status
contact switches over.
If the display flashes during operation or an error message ap-
pears, please refer to bullet "Troubleshooting".
Please refer to the data sheet for performance data and max-
imum ratings.
• no residual mois-
ture in measuring
gas / no cable break
With temperature sig-
nal option
T
= -20 °C ≙ (-4 °F) ->
Cooler
4 mA/ 2 V
T
= 5 °C ≙ (41 °F) ->
Cooler
9 mA/ 4,5 V
T
= 60 °C ≙ (140 °F) ->
Cooler
20 mA/ 10 V
Bühler Technologies GmbH
13

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis