Guidance and manufacturer's declaration – Immunity Emissions
The SUPERVEGA EVO SUCTION UNIT is intended for use in the electromagnetic environment specified below. The
customers or the user of the SUPERVEGA EVO SUCTION UNIT should make sure that it's used in such an environment.
Immunity Test
Level indicated by the
EN 60601-1-2
Electrostatic
+/-8kV on contact
discharge (ESD)
+/-15kV in air
EN 61000-4-2
Electrical fast
± 2kV power supply
transient / burst
lines
EN 61000-4-4
± 1kV for input /
output lines
Surge
± 1kV differential mode
EN 61000-4-5
+/-2 kV ordinary mode
Loss of voltage, brief
<5% U
voltage interruptions
for 0,5 cycle
and variations
40% U
EN 61000-4-11
for 5 cycle
70% U
for 25 cycle
<5% U
for 5 sec
Magnetic field with
30A/m
network frequency
(50/60 HZ)
EN 61000-4-8
Note U
is the value of the power supply voltage
T
Guidance and manufacturer's declaration – Immunity Emissions
The SUPERVEGA EVO SUCTION UNIT is intended for use in the electromagnetic environment specified below.
The customers or the user of the SUPERVEGA EVO SUCTION UNIT should make sure that it's used in such an envi-
ronment.
Immunity Test
Level indicated by the
EN 60601-1-2
Conducted
3Vrms 150kHz
Immunity
to 80Mhz
EN 61000-4-6
(for non life-supporting
devices)
Radiated
10V/m 80MHz
Immunity
to 2.7GHz
EN 61000-4-3
(for non life-supporting
devices)
ENGLISH
Compliance Level
The device doesn't
change its state
The device doesn't
change its state
The device doesn't
change its state
(>95% dip U
)
- -
T
T
(60% dip U
)
T
T
(30% dip U
)
T
T
(>95% dip U
)
T
T
The device doesn't
change its state
Compliance
Level
V1 = 3 V rms
E1 = 10 V / m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic ma-
terial, the relative humidity should be at least
30%.
Mains power quality should be that of a typical
commercial environment or hospital.
Mains power quality should be that of a typical
commercial environment or hospital.
Mains power quality should be that of a typi-
cal commercial environment or hospital If the
user of the SUPERVEGA EVO SUCTION
UNIT request that the appliance operates
continuously, the use of a continuity unit is
recommended.
The power frequency magnetic field should
be measured in the intended installation lo-
cation to make sure that it's sufficiently low.
Electromagnetic environment - guidance
The portable and mobile RF communication devices,
including cables, must not be used closer to the SU-
PERVEGA EVO SUCTION UNIT device, than the sep-
aration distance calculated by the equation applicable
to the transmitter frequency.
Recommended separation distance
3,5
√P
d=
V
1
from 80 MHz to 800MHz
12
√P
d=
E
1
from 800 MHz to 2.7 GHz
23
√P
d=
E
1
28