STD GAS AUTO-CALIBRATION KIT (OPTIONAL)

When the STD gas auto-calibration kit MCO-SG is installed it is possible to calibrate CO
density manually.
2
1. Connect a standard gas cylinder to the
Lower right side
connecting port on the lower right side of the
incubator. Use a cylinder with the same CO
2
density as the density set on the incubator.
Note: There is no problem in keeping the standard
gas cylinder connected after CO
gas density
2
calibration is completed.
Connecting port for standard gas cylinder
2. Go to the Tools #1 screen (Menu
Tools#1).
Press "STD Gas Setting" to display the STD Gas
Setting screen.
3. Enter the CO
density of the connected standard
2
CO
gas cylinder. Press "Apply". The display
2
returns to the Tools #1 screen.
Settable range: 4.0 %–21.0 %; factory setting:
5.0 %.
Note: Don't turn the CO
STD Gas Density to OFF.
2
4. Press "STD Gas Calibration" to display the STD
Gas Calibration screen.
62