Herunterladen Inhalt Inhalt Diese Seite drucken

Mizuho SUGITA Gebrauchshinweise Seite 10

Titan aneurysma-clips
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 22
allergies to metals or exogenous materials.
8) This clip must not be used in patients with
infectious diseases or sepsis as a complication.
9) This clip must not be sterilized using low-
temperature hydrogen peroxide gas plasma.
This clip is not compatible with sterilization using
low-temperature hydrogen peroxide gas plasma.
[Sterilization using such method may cause surface
color deterioration and affect its characteristics.]
10. Warning
1) It is recommended the patient is informed about the
procedure and characteristics of this clip.
2) Closing status of the blades of clip should be
verified immediately following surgery.
[The most serious adverse events that may be
reported in conjunction with this clip include
displacement of the clip, slip-out of the clip,
ejection of the clip from the clip applier, clip
fracture, laceration of a cerebral aneurysm,
-9-