4.2.5 F1.4 – Zero
4.2.6 F1.5 – Tare
30232252A ICS226
All manuals and user guides at all-guides.com
F1.3.3 – Linearity calibration
1. E SCL appears, unload scale.
Perform calibration
2. Start calibration with
3. Place the first weight on the scale and press
4. Remove the first weight and place the second weight on
donE appears, calibration is completed.
F1.4.1 – Automatic zero setting
Factory setting
0.5d
Possible settings
Off, 1d, 2d, 5d, 10d
F1.4.2 – Power up zero
2%*
Power up zero within a range of +/-2%
20%
Power up zero from -2% ~ %18
* For approved mode, the setting is always 2%.
F1.4.3 – Pushbutton zero
Factory setting
2%*
Possible settings
2%, 20%, Off
* For approved mode, the setting is always 2%.
F1.5.1 – Auto tare
Factory setting
Off
Possible settings
On, Off
F1.5.2 – Auto clear tare
Off
Factory setting
Possible settings
On, 9d
F1.5.3 – Pushbutton tare
On
Factory setting
Possible settings
Off
.
After capturing the zero point, the first weight is shown on
the display.
The first weight is confirmed. The second weight is shown
on the display.
the scale and press
.
.