9. Specifications
Specifications
Non-linearity
EN
Non-repeatability
Stability per year
Permissible temperature
ranges
Medium
Ambient
Storage
Rated temperature range
Temperature coefficients in rated temperature range (actively compensated)
Mean TC of zero
Mean TC of span
CE mark
Pressure equipment
directive
EMC directive
RoHS directive
Production environment
Packaging
Shock resistance
Vibration resistance
Electrical safety
Short-circuit resistance
Reverse polarity protection
34
Models WUD-20, WUD-25 and WUD-26
≤ 0.1 % of span (≤ 0.15 with measuring ranges ≤ 2 bar) (BFSL) per IEC 61298-2
≤ 0.12 % of span
≤ 0.25 % of span typ. (at reference conditions)
-20 ... +100 °C [-4 ... +212 °F]
-10 ... +60 °C [14 ... 140 °F]
-10 ... +60 °C [14 ... 140 °F]
-20 ... +80 °C [-4 ... +176 °F] (actively compensated)
≤ 0.1 % of span/10 K
≤ 0.15 % of span/10 K
2014/68/EU
2014/30/EU, EN 61326 emission (group 1, class B) and interference immunity
(industrial application)
2011/65/EU
Clean room class 5 per ISO 14644
Double-bagged per SEMI E49.6
15 g (11 ms); 30 g (6 ms)
7.5 mm or 2 g (1 ... 200 Hz) / 5 g
S+ vs. U-
U+ vs. U-
WIKA operating instructions Ultra High Purity transducer, model WUD-2x