Herunterladen Inhalt Inhalt Diese Seite drucken

Beurer IH 30 Gebrauchsanweisung Seite 68

Inhalator
Vorschau ausblenden Andere Handbücher für IH 30:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 4
Table 4
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment
and the IH30
The IH30 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the IH30 can help electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and
the IH30 as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2
0.12
0.38
1.2
3.8
12
68
(m)
80 MHz to 800 MHz
d = 1.2
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
d = 2.3
0.23
0.73
2.3
7.3
23

Werbung