10
Technical Specifications
(1) Model Name:
(2) Dimension and
Weight:
(3) Optical System:
(4) Focusing
Mechanism:
(5) Stage:
(6) Revolving
Nosepiece:
(7) Electrical
Specifications
• Lamp Rating:
• Average lamp
lifetime :
• Output Rating:
• Input Rating:
ECLIPSE E200 (Microscope basic unit)
227(W) ×382(D) ×415(H)mm, 10Kg
CF infinity corrected optical system
Second objective focal length f = 200 mm
Built-in diascopic illumination system (Simplified Kohler's illumination system)
Fine focus knob graduation:
Fine focus knob travel:
Coarse focus knob travel:
Stage vertical movable range:
Stroke:
X axis: 77 mm
Y axis: 53 mm
4-hole fixed type
Halogen lamp 6V-20W (PHILIPS 7388 or OSRAM HLX 64250) or Halogen lamp
6V-30W (PHILIPS 5761)
Halogen lamp 6V-20W : 100hrs.
Halogen lamp 6V-30W : 100hrs.
6V 5A max.
Model for 100, 110, and 120 V Areas
• Input voltage:
Select from 100 V, 110 V or 120 V AC by relocating
the fuse holer in the AC inlet.
• Frequency:
50/60 Hz
• Voltage fluctuation: ±10%
• Rated current:
0.8 A max.
• Fuse rating:
250 V, 1A, time-lag low-breaking type, 5×20 miniature
fuse ×2
• Power Cord:
Use only the following power supply cord. Using the
wrong power cord could result in danger or fire.
The protection Class I equipment should be connected
to PE (protective earth) terminal.
• UL listed detachable power cord set. 3-conductor
grounding type SVT, No.18 AWG, 3 m long
maximum, rated at 125 V AC minimum.
85
2µm/graduation
0.2 mm up or down / revolution
about 37.7 mm up or down / revolution
1.5 mm upward and 25 mm downward
from the focal plane.
10