The file size expected and maximum storage time for the measurement
are calculated and displayed according to the applicable parameters
selected.
The table below contains information on the triggers available:
Designation
Description
Keystroke
Measurement is started by tapping on the corre-
sponding button
Level
Measurement is started on exceeding or dropping
below a limit value
Window
Measurement is started on exceeding or dropping
below one of two limit values for a defined
measuring range
Time
Measurement is started at a defined moment of
time
External
Measurement is started on transition of a digital
signal from High to Low
Measurement is started on transition of a digital
signal from Low to High
Channel
Measurement is started when defined
warning value
warning values occurs
Channel alarm
Measurement is started when defined
value
alarm values occur
INFORMATION
Please note that the D-IN connection must be activated for the exter-
nal trigger. Refer to Chapter & "D-IN/D-OUT F1/F2".