Herunterladen Inhalt Inhalt Diese Seite drucken

WelchAllyn 88002 Bedienungsanleitung Seite 21

Video-kolposkop
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 52
5. Push the retainer spring toward the back of the colposcope and remove the lamp from its housing.
6. Remove a Welch Allyn replacement lamp # 09800 from its package. Do not touch the lamp itself or the interior
reflective surface of the lamp. Skin oils will cause premature lamp failure. Hold lamp by the outside of the reflector or
the connector only. Remove any grease or fingerprints with a clean cotton swab moistened with alcohol. Do not leave
any lint on the lamp.
7. Holding the retainer spring as in step 5, slide the new lamp into the lamp housing so that the lamp's alignment pin
engages the alignment slot in the lamp holder. (Make sure lamp is properly seated and snaps into place.)
8. Plug the electrical connector onto the new lamp.
9. Slide the lamp access door securely back into place, reversing the process described in step 3 above.
10. Place the video colposcope back into the handle, making sure the power connector is engaged.
11. Tighten the knob by rotating clockwise until snug.
12. Plug the power cord back into the electrical outlet.
18

Werbung

Inhaltsverzeichnis

Fehlerbehebung

loading

Diese Anleitung auch für:

8800488006890028900489006

Inhaltsverzeichnis