Herunterladen Inhalt Inhalt Diese Seite drucken

Supplying Power; Connecting The Transformer; Charging The Batteries; Connecting The Pump Motor And Cathode - struers Movipol-3 Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Supplying Power

Connecting the Transformer

Charging the Batteries

Connecting the Pump Motor
and Cathode

Connecting the Anode

External Etching Operations

Movipol-3
Instruction Manual
Check that the mains voltage corresponds to the voltage stated on the type
plate on the side of the machine.
 Connect the mains cable to the mains.
 Connect the 42V multi plug to the left side of Movipol-3.
 Switch O/Charge - l/Pump to top position, the batteries of
Movipol-3 are now being charged. When the batteries are fully
charged, the Charge Indicator lights green. If the batteries need
charging the Charge Indicator will light red.
Even though the batteries are factory charged, Struers recommends
the charging of Movipol-3 batteries 1-2 hours before operating the
apparatus.
A fully charged set of batteries will self-discharge in 6 months, and a
complete discharging may reduce the lifetime. Therefore the batteries
should be charged at regular intervals.
When the transformer is being used Movipol-3 should only be operated
when the Charge Indicator lights green or yellow.
 Connect the multi plug to the socket in the control unit with the
screw cap. Both the electrolyte pump motor and the cathode in
the pencil are now connected to the control unit.
 Place the banana plug of the anode wire in the socket, Anode.
The connection to the sample is established with the enclosed
magnet or clamp.
 When external etching operations should be carried out, i.e.
when using a cathode connection other than the built-in cathode
in the pencil, the black socket, Output Volt 2-90V DC should be
used as cathode connection.
IMPORTANT
WARNING
IMPORTANT
5