Mass Unit Conversion

Purpose
With this application program you
can change the weight value dis-
played from the basic weight unit to
any of 4 application weight units (see
table on next page).
Preparation
§ Select the Unit application for
toggling weight units:
see chapter on "Configuration"
(Parameter Settings)
§ Set the following parameters:
Application program
unit Toggle wt, unit
dec.plcs. Display accuracy
o All
minus 1 Reduced by 1 place
o = Factory setting
Printout for Wt. Unit Toggling
+
100.0 g
N
+ 0.22046 lb
13-Jan-2005
N
+ 3.5275 ozt
42
All decimal places
: Weight with 16-character data output format
: Weight with 22-character data output format
: Data output format for two-line printout:
08:35
: Date/time and weight
Features
– Set the basic unit and display accu-
racy in the Setup menu: see "Con-
figuration."
– Set the application weight units
and display accuracies in the
Application menu.
– These settings are stored in bat-
tery-backed memory.
– The basic unit is active when the
balance is powered up.