Invacare® HomeFill® II
INDICATOR LIGHT COLOR
Full - GREEN
ATTENTION - RED (Audible Tone)
6.2 Troubleshooting
PROBLEM
No indicator lights on when compressor
turned on.
RED light On
24
Downloaded from
www.Manualslib.com
COMPRESSOR STATUS
Compressor is On. Cylinder is finished filling
(full) - remove cylinder (FULL light is On in
>10 mins.)
or
Cylinder is not attached properly - check
connection (FULL light is On in <10 mins.).
Compressor is On. Cylinder NOT filling
Causes:
Full cylinder attached when unit turned On -
replace with empty cylinder
or
Internal failure - contact dealer.
SOLUTION
1. Turn compressor Off (O).
2. Check that power cord is properly plugged into compressor and wall outlet.
3. Turn the compressor On ( ).
4. Clean air filters. Refer to 5.2 Cleaning the Compressor Filter, page 21.
5. If performing Steps 1-4 does not cause any indicator lights to come on, contact your home
health care provider or Invacare for service.
1. Check the cylinder gauge to see if it is full. Cylinder full - Remove the cylinder and turn the unit
Off (O). Perform the 4.2 Cylinder Prefill Inspection, page 16 with a cylinder that is not full.
Cylinder not full - Proceed to step 2.
manuals search engine
EFFECT
Cylinder filling stops
Cannot fill another cylinder until compressor
is turned Off and On again
O
flowing to user if concentrator is On.
2
Compressor does not begin to fill or stops
filling
Unit MUST be turned Off and On before
filling can occur.
O
flowing to user if concentrator is On.
2
1145804-E