Herunterladen Inhalt Inhalt Diese Seite drucken

Appendix 1: Fm-Csa Installation Drawing - WIKA T53.10 Betriebsanleitung

Feldbus temperatur-transmitter
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Appendix 1: FM-CSA Installation Drawing

Model T53.10.0IS
Hazardous (classified) Location
Class I, Division 1, Groups A, B, C, D
or
Class I, Zone 0, IIC
Approved
Termina-
tion
Sensor
Terminal 3, 4, 5, 6
V
or U
: 5.71 V
t
o
l
or l
:
8.4 mA
t
o
P
or P
: 12 mW
t
o
C
or C
: 40 µF
a
o
L
or L
: 200 mH
a
o
Class I, Zone 0, Ex ia IIC, Entity / FISCO
IS, Class I, Division 1, Group A, B, C, D Entity / FISCO
Barrier type Linear barrier Trapezoid
T1 ... T4
T a ≤ +85 °C T a ≤ +75 °C T a ≤ +85 °C T a ≤ +85 °C
T5
T a ≤ +70 °C T a ≤ +65 °C T a ≤ +60 °C T a ≤ +60 °C
T6
T a ≤ +60 °C T a ≤ +45 °C T a ≤ +45 °C T a ≤ +45 °C
V
or Ui
30 V
max
l
or li
120 mA
max
Pi
0.84 W
Ci
2.0 nF
Li
1 µH
See Installation notes.
Installation Drawing 11175631.01
WIKA Operating Instructions Temperature Transmitter T53
Sensor
Sensor
Terminal 1, 2
Suitable
barrier
for FISCO
systems
30 V
17.5 V
300 mA
250 mA
1.3 W
2.0 W
2.0 nF
2.0 nF
1 µH
1 µH
Unclassified Location
Associated Apparatus
Barrier or
FISCO Supply
with
entity Parameters:
UM ≤ 250 V
V
or U
≤ V
or U
oc
o
max
l
or l
≤ l
or l
sc
o
max
i
P
≤ P
o
i
C
or C
≥ C
+ C
a
o
i
cable
L
or L
≥ L
or L
a
o
i
cable
The device must not be
connected to any
associated apparatus
which uses or generates
more than 250 VRMS
Suitable
for FISCO
systems
15 V
900 mA
5.32 W
2.0 nF
1 µH
GB
i
1/6
25

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis