Specifications in relation to the user
User's interface
Display
Display :
Current output
Relay state
Programming
Protection
Specifications in relation to the environment
Ambient conditions
Storing and operating temperature
Humidity
Enclosure protection rating
Sensor 8170 protection rating
Construction
Dimensions maximum
Weight
Materials
Electronic housing
Front foil
Sensor 8170
Conformity to standards
Emission
Immunity
Safety
8175
SPECIFICATIONS
8 digits alphanumeric LCD 9mm high
Generated current indication : xx.xx mA
Red LED's on when contact is closed
Menus with 3 programming keys
Lockable switch for the 'Enter' key
-20 to +60°C / -4 to +140°F
-40 to +80°C / -40 to +176°F (for the remote sensor)
Maximum of 80%, non condensated
IP65 (Compact, Wall & Panel front)
IP20 (Interior of panel version)
IP67
See section 6.3
1kg maximum
PC (glass reinforced fiber) (compact and
panel-mounted versions)
ABS (Wall version)
Polyester
PVDF /POM
According to generic norm EN 50081.1
According to generic norm EN 50082.2
According to safety regulations for measuring instruments
for regulation and laboratory NF EN 61010-1