1. Error messages

Failure encountered
Possible cause
Display
Excessive cuff pressure.
Err - 300
Cuff inflated to maximum
due to movement of the
arm or body.
Display
Measurement error;
Err - 1
measurement could not
be carried out due to
movement or speaking.
Display
Inflation error, pressure
Err - 2
built up too quickly or too
slowly.
Cuff is not correctly fitted.
Display
Deflation rate too high or
Err - 3
too low.
Movement or talking
during measurement.
Display 0
Start/Stop button acci-
0
dentally pressed while
batteries were being
changed.
Irregular heartbeat, dis-
ruptive movements, shak-
ing, wobbling (objects),
breathing in deeply, etc.
Measured values/
Too little rest before the
pulse values are ex-
measurement. Movement
tremely high or low.
or talking during meas-
urement.
Incorrect measurement
posture
Consecutive meas-
Blood pressure is not a
urements produce
fixed quantity. It may vary
different measured
in the upward or down-
values (readings).
ward direction within a
few minutes by more than
20 mmHg.
EN-46
Technical information
Corrective action
- Repeat the measurement
- Do not move your arm
- Do not speak
- Repeat the measurement
- Do not move your arm
- Do not speak
- Repeat the measurement
- Do not move your arm
- Do not speak
- Check wrist circumference.
- Repeat the measurement
- Do not move your arm
- Do not speak
Switch unit off and then on
again with Start/Stop button.
Repeat measurement after 3-5
minutes rest. Possible influenc-
es on the measurement results
(depending on the severity
of the arrhythmia) must be
discussed with your doctor!
Repeat the measurement after a
break of 3-5 minutes. Pay atten-
tion to the usage information
on page 36.
Repeat the measurement and
maintain the correct posture.
(Page 40)
In order to obtain comparable
results, always carry out the
measurements under the same
conditions.
Technical information
Failure encountered
Possible cause
Measurement was
Device detects a fault or
interrupted, the cuff
that the diastolic value is
deflated to 0 and
too low.
the measurement
Movement during meas-
restarted.
urement
No display after unit
Batteries inserted incor-
is switched on
rectly.
Batteries flat.
Battery contacts dirty.
Nothing appears
No measured values have
in the display after
been stored.
pressing the Memory
button
Measurement inter-
Batteries flat.
rupted
The
symbol
Low batteries
appears during
measurement but
the measurement
continues.
Date/time not shown
Function accidentally
in display
deactivated

2. Customer service

Device repairs may only be performed by the manufacturer or an expressly
authorised body. Please contact:
UEBE Medical GmbH
Bürgermeister-Kuhn-Straße 22
97900 Külsheim, Germany
info@uebe.com
www.uebe.com
E
Corrective action
Device shows no fault. Repeat
measurement
If necessary, interrupt meas-
urement, measure again after a
break of 5 minutes.
Check position of batteries.
Change batteries.
Clean battery contacts with a
dry cloth.
Repeat measurement
Change batteries.
You can still perform approx. 30
measurements, but should keep
spare batteries available.