Cautions
Use only compatible accessories described in these
Instructions for Use (IFU).
Do not sterilize the M2741A sensor or immerse it in
liquids.
Do not apply any tension to the M2741A sensor cable.
Do not exceed storage or operating temperature limits
(see limits under Product Symbols earlier in this IFU).
Observe precautions for electrostatic discharge (ESD)
and electromagnetic interference (EMI) to and from
other equipment.
Disconnect the cannula, airway adapter or sample line
from the sensor when not in use.
Troubleshooting tips
Check the following whenever a change in patient CO
waveform is observed:
Check patient status.
Check positioning of endotracheal tube or nasal cannula.
Check that airway adapter is not damaged and is securely
connected to ventilator circuit.
If problem is not identified, change the Airway Adapter
or cannula.
If problem is still not identified, change the M2741A
sensor.