Cautions
•
Use only compatible accessories described in these
Instructions for Use (IFU).
•
Do not sterilize the M2741A sensor or immerse it in
liquids.
•
Do not apply any tension to the M2741A sensor cable.
•
Do not exceed storage or operating temperature limits
(see limits under Product Symbols earlier in this IFU).
•
Observe precautions for electrostatic discharge (ESD)
and electromagnetic interference (EMI) to and from
other equipment.
•
Disconnect the cannula, airway adapter or sample line
from the sensor when not in use.
Troubleshooting tips
Check the following whenever a change in patient CO
waveform is observed:
•
Check patient status.
•
Check positioning of endotracheal tube or nasal cannula.
•
Check that airway adapter is not damaged and is securely
connected to ventilator circuit.
•
If problem is not identified, change the Airway Adapter
or cannula.
•
If problem is still not identified, change the M2741A
sensor.