®
R&S
ZVA-Z75, -Z90, -Z110, -Z140, -Z170, -Z220, -Z325, -Z500
Figure 3-1: Frequency Converter dialog (example: R&S ZVA-Z110)
3.4 Connecting the Frequency Converters
Each frequency converter must be connected to the analyzer, the power supply and
the DUT. Please refer to the following sections for details.
●
Analyzer ports:
●
Power supply:
on page 15
●
DUT (usually connected after calibration):
on page 16
3.5 Calibration
The R&S ZVA-Z90 and the latest variant of the R&S ZVA-Z110 (order number
1307.7000.03) allow for a variation of the output power by changing the input power.
With these converter types the "R&S ZVA Frequency Converter Leveling Tool" can be
applied for the linearization of output power in the frequency range of interest. See
documentation "Software Utility R&S ZVA frequency converter leveling tool. Getting
Started" for further details.
With other converter models, the output power can only be manually set using the
adjusting screw. This does not allow for a power flatness calibration. A power calibra-
Getting Started 1307.7039.62 ─ 11
Chapter 2.4.5, "Connecting RF
Chapter 2.4.6, "Connecting the Converter to the DC
Cables",
on page 12
Chapter 2.4.8, "Mounting a
Basic Operation
Calibration
Supply",
DUT",
20