ALARM AND WARNING MESSAGES
PRIMARY SECONDARY
DISPLAYS
OCCL
INTO PUMP
OCCL
OUT OF PUMP
ATTN
PUMP ON HOLD
BATT
LOW BATTERY
BATT
BATT EMPTY
DOOR
OPEN
NO
CASSETTE
Page 44
CONDITION OR
PROBLEM
Flow has stopped
due to occlusion
between the
formula container
and the pump.
Flow has stopped
due to occlusion
between the
pump and the
patient.
Hold time has
elapsed.
Pump battery
is nearly empty.
Approximately
30 minutes of
battery operation
remain.
Pump battery is
depleted. Pump
has stopped
feeding.
The door is open
while feeding
or priming the
feeding set.
Cassette is
not installed or
was installed
incorrectly when
user attempted
to run pump or
prime the feeding
set.
EXPLANATION OR CORRECTIVE ACTION
•
Turn dial to HOLD
•
Remove feeding set from pump. Clear
obstruction; check for kinked tubing or
empty container. Reinstall feeding set
•
If this alarm appeared during priming,
attempt to prime the feeding set again,
then connect the feeding adapter to your
enteral feeding tube
•
Turn dial to RUN to begin feeding
•
Turn dial to RUN to begin feeding
•
Connect the pump to live AC/Mains
power.
•
Connect the pump to live AC/Mains
power. The display will indicate "CHRG
PLEASE WAIT"
•
Turn dial to HOLD
•
Close the door
•
Turn dial to RUN to begin feeding, or
attempt to prime the feeding set again
•
Turn dial to HOLD
•
Insert cassette and close door
•
Turn dial to RUN to begin feeding, or
attempt to prime the feeding set again