Herunterladen Inhalt Inhalt Diese Seite drucken

Replacing The Rechargeable Battery - Dräger X-act 5000 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für X-act 5000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 28
When the rechargeable battery is charged while inside the instrument,
the battery icon will be indicated on the display.
When the rechargeable battery is charged separately, the LED (2) will
light red during charging. The LED (2) will light green when the
rechargeable battery is completely charged. If the LED (2) blinks red
during charging, an error occurred.
If the LED (2) blinks red during charging, an error occurred. To clear
the error proceed as follows:
1. Disconnect the rechargeable battery from the charger for approx.
10 seconds.
2. Reconnect the charger.
3. If the LED (2) lights red, continue the charging procedure.
If the LED (2) alternates between green and red the rechargeable
battery is deeply discharged. In this case, proceed as follows:
1. Leave the rechargeable battery on the charger until the LED (2)
blinks red.
2. Disconnect the rechargeable battery from the charger for approx.
10 seconds.
3. Reconnect the charger.
4. If the LED (2) lights red, continue the charging procedure.
Dräger X-act 5000
3.4.2

Replacing the rechargeable battery

WARNING
!
Do not remove or replace the rechargeable battery in areas
subject to possible explosion.
1. Remove the 2 screws with an appropriate hex key.
2. Pull the rechargeable battery out of the Dräger X-act 5000.
3. Insert a charged NiMH battery into the Dräger X-act 5000.
4. Fasten the 2 screws with an appropriate hex key.
Use
00721900.eps
19

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis