Cleaning

NOTE on material damage
− Do not immerse the scale in water or other liquids.
− Do not under any circumstances use harsh, scouring or abrasive
cleaning agents or the like to clean the scale. Otherwise, the surface
could be scratched.
− Wipe the scale as needed with a slightly damp soft cloth.
− Disinfect the weighing surface and the sensors from time to time using a suitable
disinfecting agent (available at specialist retailers and pharmacies).

Fault/Remedy

Fault
No function
The scale appears to be displaying
wrong weighing results
"Err" is shown on the display.
The display shows "Er".
"Lo" is shown on the display
Body Analyser Scales BA 1807
Possible cause
‚ Is the battery empty or inserted the wrong
way around?
‚ Is the scale located on a firm, even surface?
‚ Is a device that can cause electromagnetic
disturbances located in the vicinity of the
scale?
‚ Remove the battery from the scales and
wait a moment. Reinsert the battery. The
data stored up to this point will be retained.
‚ The maximum capacity of the scale (180 kg)
has been exceeded.
‚ You have left the scale during the ongoing
body analysis.
‚ The proportion of fat in % is too low or too
high and can therefore not be displayed.
‚ The battery is empty (see "Changing the
battery").
Cleaning
27